The ABA option allows continuous, SLOW acquisition of the Inputs, before, during and after a trigger event. At the trigger event the inputs can be sampled with the programmed FAST sampling rate.
•Combination of a slow data logger with a fast digitizer.
•Similar to Multiple Recording option, but with data in between the segments
•In combination with the timestamp option this option gives a complete acquisition with a dual-timebase
•The timestamp will mark the beginning of the second (faster) B-timebase
•Can be used without timestamps for monitoring purposes
Data is stored within a separate FIFO on the card, no space of the on-board memory is lost. Data buffers for the two timebases are completely separated and will not influence each other.